VC / NIC642

← Infant Radiant Warmer Open Care System Apple SeriesIn catalogue
Infant Radiant Warmer Stand with Microprocessor based temperature controller with 4 mode Skin / Air / Manual / Prewarm - NIC643
Technical Specification: Heater Box Consists of Ceramic Infra Red Heater 650Watts, The source box can be swivelled to take X-ray Due to temp. insulated cover on the heater box, its exterior surface remains cool even after long run The construction of the stainless steel reflector and Guard rail is such that the complete cleaning of heater & reflector is possible without any electrical shock hazards Halogen Examinati
Enquire about this product ↗At a glance
Clear specs.
Better decisions.
Use these core specifications as a guide. Contact our team for full technical sheets, options and regional availability.
Feature 1Consists of Ceramic Infra Red Heater 650Watts,
Feature 2The source box can be swivelled to take X-ray
Feature 3Due to temp. insulated cover on the heater box, its exterior surface remains cool even after long run
Feature 4The construction of the stainless steel reflector and Guard rail is such that the complete cleaning of heater & reflector is possible without any
electrical shock hazards
Feature 5To Provide accurate assessment of Infant colour and sufficient illumination during procedures
Feature 6Clear acrylic collapsible side panels
Feature 7Stainless steel baby tray with transparent slot to incorporate the facility of under surface phototherapy
Feature 8Acrylic tray and cushion mattress with head up/down facility
Feature 9Acrylic X-ray tray with X-ray guide for positioning the cassette
Feature 10The design aspects also allow ease of dismantling baby care area during cleaning or fumigation
Feature 11Side Mounted Monitor tray, Pole for Syringe Pump / Infusion Pump
Feature 12S.S. IV Stand.
Feature 13To Prove Highest Degree of Accuracy of maintenance of baby’s body Temperature.
Thermistor type sensor with Accuracy± 0.2°C
Feature 15Audio alarm mute facility.
Feature 16Mute timing adjustable
Skin Mode30.0°C to 38.0°C
Air ModeRoom temp. 20.0°C to 39.0°C
Manual ModeVary Heater Power % from 0% to 100% with increment of 5%
Time duration (Manual mode)30 mins (factory default)
Prewarm modeVary Heater Power % from 0% to 25% with increment of 5%
Feature 231” Bright 7 segment LED’s (3 nos.) for Skin temp.
Feature 240.5” Bright 7 segment LED’s (3 nos.) for Set temp.
Feature 250.5” Bright 7 segment LED’s (3 nos.) for Heater Power%.
Feature 260.5” Bright 7 segment LED’s (3 nos.) for Air temp.
Resolution0.1°C
Feature 28Control Mode
Feature 29Constant Skin temp in ° Fahrenheit.
TimerDown Timer / APGAR Timer
Alarm Messages for Skin Over temperatureTemp > 38.0°C Skin Under temperature: Temp < 34.0°C Skin Positive band: Temp > Set Temp by 1.0°C Skin Negative Band: Temp < Set Temp by 1.0°C Air Over temperature: Temp > 39.0°C Air Under temperature: Temp < 20.0°C Air Positive band: Temp > Set Temp by 1.5°C Air Negative Band: Temp > Set Temp by 3.0°C Heater Failure: Heater connection is failed/damage/faulty System Failure: Fluctuation of voltage supply. Check Sensor position: No proper voltage supply
Skin Over temperatureTemp > 38.0°C
Skin Under temperatureTemp < 34.0°C
Skin Positive bandTemp > Set Temp by 1.0°C
Skin Negative BandTemp < Set Temp by 1.0°C
Air Over temperatureTemp > 39.0°C
Air Under temperatureTemp < 20.0°C
Air Positive bandTemp > Set Temp by 1.5°C
Air Negative BandTemp > Set Temp by 3.0°C
Heater FailureHeater connection is failed/damage/faulty
System FailureFluctuation of voltage supply.
Check Sensor positionNo proper voltage supply
Ready to specify NIC642?